Evaluation Of A New Direct-Comparison Aniseikonia Test                   View Shopping Cart

(Analysis based on Version 2 of the Aniseikonia Software)

Gerard C de Wit PhD

The following technical paper to appear in as follows:  Evaluation of a new direct-comparison Aniseikonia test, Bin. Vis. & Strabismus Q. 18: 87-94 (2003)

Note: Tables and Figures may come up slowly due to file sizes

ABSTRACT

Background & Purpose: Aniseikonia is a condition in which the two eyes perceive images of different size or shape, causing a variety of visual symptoms including asthenopia. Besides anisometropes (with a prevalence of 5-10% in the population above age 20 years), also pseudophakes and refractive surgery patents are at risk. For example, 40% of the pseudophakes seem to suffer from Aniseikonia. Reliable measurement and management of Aniseikonia is therefore important. The "Aniseikonia Inspector" is a new, commercially available, software product to measure an manage Aniseikonia. The purpose of this study is to evaluate this Aniseikonia test of the Aniseikonia Inspector.

Methods: Aniseikonia was induced in four subjects, with normal vision, by means of afocal size lenses.  Using the Aniseikonia Inspector, the resulting Aniseikonia was measured in vertical, horizontal and diagonal directions.

Results: The average ratio between the measured Aniseikonia and the induced Aniseikonia was 0.98, 0.89 and 0.93 respectively for the vertical, horizontal and diagonal directions. For two consecutive measurements. Of the same aniseikonic state, the difference in measurement value was 97%, 75% and 94% of the time within one resolution step size (0.5% horizontally and vertically, and 0.7% diagonally). 

Conclusion: Aniseikonia was measured accurately. Measurements in the vertical direction were more accurate that in the diagonal or horizontal directions, which is probably due to fixation disparities. The Aniseikonia Inspector is a very useful new tool in treating the growing number of Aniseikonia patients. 

Key Words: Aniseikonia / anisometropia / binocular vision / eikonometry / fixation disparity / test

* The author and publisher have a commercial interest in The Aniseikonia Inspector used in this study.

INTRODUCTION

Incidence of Aniseikonia

After an initial burst of interest in Aniseikonia in the 1950's, the condition seems to be out of vogue today (1,2). In contradiction, the number of patients suffering from Aniseikonia have increased considerably. Not only anisometropes (with a prevalence of 5-10% of the population above the age of 20 years (3)) and other people with inherent optical or anatomical differences between the two eyes are at risk, but now also people who have had refractive surgery (e.g. LASIK, PRK, etc.) and cataract surgery are at risk (4,5). For example, Kramer et al. (5) found that 40% of all pseudophakes have ophthalmic complaints referable to Aniseikonia. Nowadays, in the United States alone, more than 1.5 million cataract operations are performed each year, which means that Aniseikonia can be considered a significant public health issue (6,7).

Aniseikonia Management

The question remains why eye care professionals pay relatively little attention to Aniseikonia today. This will be discussed after considering the basic steps of Aniseikonia management (Link to Figure 1)


Measure Aniseikonia

Verify if patient is helped by correcting Aniseikonia

Determine an Aniseikonia corrected prescription

The first step "Measure Aniseikonia" may seem trivial, but because of a deficiency of accurate and good instrumentation, measuring Aniseikonia has been difficult. Because of this lack, very few eye care professionals are able to even diagnose or measure Aniseikonia at all. Others have recommended one use certain "rules of thumb", for estimating Aniseikonia.  Research, however, shows that these rules of thumb (including Knapp's law) have a large error rate (8-10). 

For certain groups of patients such as pseudophakes and strabismus patients the error rate of rules of thumb was as high as 60-70% (8). So there actually was no test or method available to measure Aniseikonia, while it is important to use a reliable measurement method to manage Aniseikonia.

Verifying if a patient is helped by correcting the Aniseikonia (subjective eikonometry), often will be the second step in managing Aniseikonia. It gives the opportunity to refine the correction to maximal patient satisfaction. The tolerance level to Aniseikonia also differs a lot from patient to patient. Some patients can be really grateful when correcting 1% of Aniseikonia, while another might not be bothered by as much as 3% Aniseikonia. A common way of verification is to have the patient look through a focal size lenses with similar magnification as the Aniseikonia found in the measurements. Another way is to simulate and manipulate the Aniseikonia correction by haploscopic image projection on, for example, a computer screen.

The last step of Aniseikonia management is the calculation of an aniseikonic corrected prescription. By changing the shape factor (base curve and lens thickness), refractive index, and/or vertex distance of  spectacle lenses, the difference in magnification between the lenses can be modified to correct for the  Aniseikonia. To do this calculation by hand can be quite a challenge; this is yet another reason why eye care professionals can be hesitant to deal with Aniseikonia. However, with the advent of computers this problem can be easily solved.

A major reason, certainly, why, currently, eye care professionals pay relatively little attention to Aniseikonia is that they cannot perform one (or more) of these basic steps of Aniseikonia management or they consider it is too time consuming or unprofitable to undertake.

The Aniseikonia Inspector

Recently, a new Aniseikonia product (Richmond Products p/n 4544) has become commercially available. This software product performs and manages all three basic steps of Aniseikonia management (Fig. 1). Therefore, this product intends to make it possible and practical to give Aniseikonia the attention it deserves.

The Aniseikonia test of the Aniseikonia Inspector is based on the simple technique of direct comparison eikonometry (11). Although the technique seems remarkably uncomplicated, it can have a sensitivity of less than 0.5%. However, a good layout of the test is important to avoid underestimation of Aniseikonia. Ogle and Ames (11) mention that fusion stimuli from objects or contours surrounding the comparison targets should be eliminated. McCormack et al. (7) found that another direct comparison Aniseikonia test (the New Aniseikonia Test) significantly underestimated Aniseikonia.  They speculated that this underestimation is due to a sensory fusion response resulting in a  rescaling of the perceived image. McCormack et al. also demonstrated a direct comparison test on a computer screen that did not have a significant underestimation. The Aniseikonia test of the Aniseikonia Inspector is such a test on a computer screen and this paper evaluates the accuracy of this test.

Subjects and Methods

Subjects

Four subjects, aged 28-34 years, with normal visual acuity and a negative ocular and vision history, participated in this study to evaluate the Aniseikonia test of the Aniseikonia Inspector. 

Aniseikonia was induced in these subjects by positioning afocal size lenses (MultiLens, Sweden) in front of the right eye in random order. The thickness of the size lenses could have given the subject a slight clue about the magnification of the lens; however, the Aniseikonia test did not reveal what Aniseikonia setting was being made, so it is unlikely that this would have given a bias.

Table 1 shows the curvatures (obtained from the manufacturer) and thicknesses (measured) of these 4 lenses. Each lens was used both to magnify and to minify the OD image (by flipping the lens). The magnification values of each lens was determined by means of optical ray-tracing using the testing dimensions during the measurements in this study: 75 cm for the image distance, 16 cm image size, and 25 mm for the vertex distance. Both the magnification related to static Aniseikonia (classical Aniseikonia, i.e. difference in retinal image size with a fixed gaze direction) was calculated as well as the magnification related to dynamic Aniseikonia (optical anisophoria, i.e. unequal phorias in the various directions of gaze also due to anisometropic spectacles) (12-14). This last type of Aniseikonia is probably more appropriate for the Aniseikonia test of the Aniseikonia Inspector, because the subjects' gaze direction was not fixed during the test.

Aniseikonia Test

Figure 2 shows a (gray-scaled) image of the Aniseikonia test of the Aniseikonia Inspector. The patient looks at this screen through red-green glasses to separate and isolate binocular vision. In this study, we used red in front of OD and the dimensions were the same as those stated for the determination of the magnification values of each size lens. An Aniseikonia setting was made by changing the size of one of the two half circles with keys on the keyboard or with the mouse until the two half circles appear equal in size.  The Aniseikonia is measured in the vertical, the horizontal, and the diagonal direction. In each direction, the measurement is done twice, once starting with –25% preset Aniseikonia and once starting with +25% preset Aniseikonia. The average of these two measurements is taken as the Aniseikonia value. To verify if the Aniseikonia Inspector measured Aniseikonia correctly, the measured Aniseikonia are plotted against the (dynamic) magnifications induced by the size lenses. The ideal test result would be one where a magnification (M) results in a measured Aniseikonia (A) of: Formula Around M=0% the aniseikonic function A(M) is predominantly a linear function with a slope close to –

1. Therefore, a linear regression analysis is performed on the measured data and the slope is compared to –1.0 (or –0.995: the slope obtained when doing a linear regression analysis on Eq. (1) with, as data points, the same magnifications as in the experiments). In other words, the ideal Aniseikonia test should measure –1% of Aniseikonia for each percent of induced magnification. 

RESULTS

Figure 3 shows the measured Aniseikonia as a function of induced magnification for the vertical, the horizontal, and the diagonal directions. The slope and R2 value of the linear regression lines of the data of all subjects is shown in Table 2 .

In the vertical direction the average slope of the best fit linear regression lines is –0.98, very close to –1.0. 

This suggests that the induced Aniseikonia is measured correctly (i.e. accurately).  In the horizontal direction there may be a minor underestimation (average slope = -0.89). All of the subjects reported that inequality in this direction was more difficult to assess than the vertical direction, because the images were less stable. This also shows in the R2 values. In the diagonal direction the slope is closer to –1.0, but still there seems to be a slight underestimation of the Aniseikonia.

Table 3 shows the Y-intercept values of the linear regression lines for the different subjects, which should represent the inherent Aniseikonia of the subjects. Note that the ideal Aniseikonia curve, given by Equation 1, is not a linear function and a regression line through the ideal data points would have given a bias of 0.25%.  This bias has been subtracted from the actual Y-intercept, giving the presented 'corrected Y-intercept'.

The actual Aniseikonia measurements (i.e. no size lens present) of the subjects is also shown. This data shows that the subjects did not have a clinically significant  amount (i.e. >1%) (15) of inherent Aniseikonia.   To evaluate if the starting Aniseikonia value of a measurement had any effect on the final Aniseikonia setting, Figure 4 shows the number of occurrences of the difference between the first and second measurement of all 108 trials (4 subjects × 9 induced Aniseikonia values × 3 directions). The first measurement (M1) of each trial started at an Aniseikonia value of –25%, while the second measurement (M2) started at +25%. If the starting Aniseikonia value would have given a bias to the measurement, the distribution plots in Figure 4 would not have been symmetric around M1-M2 = 0%.  However, on average the difference between the two measurements is –0.08%, 0.01%, and –0.18% respectively for the vertical, horizontal, and diagonal direction. These values are all close to 0% and small compared to the resolution step size of 0.5% in the vertical and horizontal direction and 0.7% in the diagonal direction.

Based on the spread in values, Figure 4 also shows that the repeatability for two consecutive measurements (even thought the starting value was different) was very good. The percentage of trials in which two consecutive measurements were within one resolution step size from each other is 97%, 75%, and 94% for the vertical, horizontal, and diagonal direction.

DISCUSSION

The Aniseikonia measured with the Aniseikonia Inspector showed a very good correspondence with the induced Aniseikonia in the subjects. In the vertical direction, the slope of the linear regression line for the 4 subjects ranged from –0.95 to –1.02, so all very close to the ideal –1.0.  In comparison, McCormack et al. (7) found an average slope of -0.32 for the New Aniseikonia Test, indicating a significant underestimation. They speculated that the reason for the underestimation was the visibility of fusion stimuli from objects, texture, or contours surrounding the comparison targets. One inherent advantage of a direct comparison test on a computer screen is that the test images are themselves often brighter than the surroundings, while a test on paper requires lighting that will also illuminate and cast shadows on the surroundings.

In the horizontal direction the slope of the linear regression line for the Aniseikonia Inspector data varied more (-0.81 to –1.02) than in the vertical direction. Also the repeatability was worse than in the  vertical direction. A likely reason is that fixation disparities and phorias cause the images to be less stable (11). With the Aniseikonia Inspector it is possible to correct for these effects by displacing the two half circles relative to each other. This way the two half circles can be positioned symmetrically on top of each other for a more accurate size comparison. However, some remaining vergence interference may still cause instability in the images resulting in less reliable measurements than in the vertical direction. Ogle et al. (16) reported that the reliability of the horizontal direction depends on the viewing distance. At distant vision it is more reliable than at near vision. Ogle (11) also reported that comparison targets in the oblique meridians have proven unreliable because of the disturbing effect of fixation disparity. However, the subjects in this investigation all found the diagonal direction clearly easier to do than the horizontal direction. This is also visible in the accuracy and repeatability of the measurement data. The explanation for this incongruity might be that with the Aniseikonia Inspector it is possibility to move the half circles relative to each other to compensate for fixation disparity.

Since, in some patients, the measurement in the vertical direction might be more accurate than the measurement in the other directions, the Aniseikonia Inspector also provides an aniseikonic ellipse (i.e. complete description of the Aniseikonia (2,11)) based only on the measurement in the vertical direction plus the prescription of the spectacle glasses used during the test. The idea behind this is that, in practice, the meridional effects are nearly always produced by spectacle cylinders (12). If this is assumed, only one measurement direction is needed to calculate the overall Aniseikonia part of the aniseikonic ellipse, which is most important part of the Aniseikonia for clinical purposes (12).

CONCLUSION

The data in this paper indicates that the Aniseikonia test of the Aniseikonia Inspector measures Aniseikonia correctly and accurately, especially in the vertical direction. In the horizontal and diagonal direction, there may be a minor underestimation of the Aniseikonia. Because the Aniseikonia Inspector provides the means not only to easily measure Aniseikonia, but also to carry out the two other basic steps of Aniseikonia management (verification and prescription calculation), it presents and provides a very useful new tool in treating the growing number of Aniseikonia patients.

REFERENCES

  1. Romano PE. Aniseikonia? YECH! Binocul Vis Strabismus Q 1999; 14: 173-176.

  2. Bennett AG, Rabbetts RB. Clinical Visual Optics. 3rd ed.

  3. Butterworth-Heinemann Ltd, Oxford 1998; 273

  4. Weale RA. On the age-related prevalence of anisometropia. Ophthalmic Res. 2002; 34:389-392

  5. Lubkin V. Aniseikonia at the Millenium. Binocul Vis, Strabismus Q 1999; 14: 179-182

  6. Kramer PW, Lubkin V, Pavlica M, Covin R. Symptomatic Aniseikonia in Unilateral and Bilateral Pseudophakia. A   Projection Space Eikonometer Study. Binocul Vis Strabismus Q 1999; 14: 183-190

  7. Foster A. Vision 2020: The cataract challenge. J. of Community Eye Health 2000; 34: 1                                       (http://www.jceh.co.uk/journal/34_1.asp)    

  8. McCormack G, Peli E, Stone P. Differences in Tests of Aniseikonia. Invest Ophthalmol Vis Sci 1992; 33:2063-2067

  9. Lubkin V, Shippman S, Bennett G. et. al. Aniseikonia Quantification: Error Rate of Rule of Thumb Estimation. Binocul Vis Strabismus Q 1999; 14: 191-196

  10. Kramer P. Shippman S. Bennett G. et. al. A study of Aniseikonia and Knapp's Law Using a Projection Space Eikonometer. Binocul Vis Strabismus Q 1999; 14: 197-201    

  11. Romano PE, Von Noorden GK. Knapp's Law and Unilateral Axial High Myopia. Binocul Vis Strab. Q 1999; 14: 215-222

  12. Ogle KN. Researches in Binocular Vision. W.B. Saunders Company, Philadelphia 1950

  13. Remole A, Robertson KM. Aniseikonia and Anisophoria: current concepts and clinical applications. Runestone Publishing, Waterloo, Canada, 1996

  14. Friedenwald JS. Diagnosis and treatment of anisophoria. Arch. Ophth. 1936; 15: 283-307

  15. Scheiman M, Wick B. Clinical Management of Binocular Vision: Heterophoric, Accommodative, and Eye Movement Disorders, JB Lippencott Co. Philadelphia, 1994

  16. Michaels DD, Visual Optics and Refraction: A Clinical Approach, M Mosby, St. Louis, 1985

  17. Ogle KN, Imus HA, Madigan LF, et. al. Repeatability of Ophthalmo-eikonometer Measurements. Arch. Ophth. 1940; 24: 1179-1189

Aniseikonia Evaluation Software

Aniseikonia And Cyclodeviation Index

 

All Products In Alphabetical Order

Richmond Products Home Page                                                                                                                                      View Shopping Cart